- Model
- POC5
- Oxygen flow settings
- Pulse settings 1–5; not continuous-flow rates in L/min
- Oxygen flow volume
- Setting 1: 210 mL/min; setting 2: 420 mL/min; setting 3: 630 mL/min; setting 4: 840 mL/min; setting 5: 1,000 mL/min
- Oxygen concentration
- ≥90% (V/V)
- Maximum outlet pressure
- ≤199.3 kPa
- Product weight
- 4.85 lb / 2.2 kg including 8-cell battery; 5.73 lb / 2.6 kg including 16-cell battery
- Dimensions
- 7.3 × 3.3 × 8.2 in / 18.6 × 8.4 × 20.8 cm including 8-cell battery; 7.3 × 3.3 × 9.0 in / 18.6 × 8.4 × 22.8 cm including 16-cell battery. Dimension order as supplied; axes not specified.
- Battery options
- 8-cell or 16-cell
- Battery duration — 8-cell
- Up to: setting 1, 6 h 30 min; setting 2, 5 h; setting 3, 3 h; setting 4, 2 h; setting 5, 1 h 45 min
- Battery duration — 16-cell
- Up to: setting 1, 13 h; setting 2, 10 h; setting 3, 6 h; setting 4, 4 h; setting 5, 3 h 30 min
- Ingress protection
- IP22
- Sound level
- ≤60 dBA according to ISO 80601-2-69; <40 dB(A), typical at setting 2 (MDS-Hi; source condition not defined)
- Maximum power consumption
- 100 W
- AC input
- 100–240 VAC, 50–60 Hz
- DC input
- 13.5–15.0 VDC
- Operating temperature
- 41 to 104°F (5 to 40°C)
- Operating humidity
- 0–90%, non-condensing
- Operating altitude
- 0–10,000 ft (0–3,048 m)
- Storage and transport temperature
- −13 to 158°F (−25 to 70°C)
- Storage and transport humidity
- 0–95%, non-condensing
- Storage and transport altitude
- 0–10,000 ft (0–3,048 m)